Pressure Support<br />

This control establishes the target pressure above the PEEP setting for Pressure Support patient<br />

breaths. If Pressure Support is set to dashes “- -“, all patient breaths will be given as Spontaneous<br />

breaths. Inspiratory flow for Pressure Support and Spontaneous breaths is controlled to meet the<br />

patient demand.<br />

NOTE<br />

Pressure Control and Pressure Support breaths do compensate for PEEP. Delivered pressure is<br />

controlled by the Pressure Support setting and is affected by the PEEP setting. For example, a<br />

Pressure Control setting of 20cmH2O and a PEEP setting of 10cmH2O results in a Peak Inspiratory<br />

Pressure (PIP) of 30cmH2O.<br />

REMARQUE<br />

Le contrôle de la pression et la pression de support des respirations compensent la PEP. La pression<br />

administrée est contrôlée par le réglage du contrôle de la pression et elle dépend du réglage PEP.<br />

Par exemple, un réglage du contrôle de la pression de 20 cmH2O et un réglage PEP de 10 cmH2O<br />

donnent une pression inspiratoire maximale (PImax) de 30 cmH2O.<br />

To set Pressure Support:<br />

1) Push the Pres. Support button.<br />

2) Change the setting using the Set Value knob.<br />

Range: “- -”, 1 - 60 cmH2O<br />

Pressure Support breaths may be terminated by flow or by time.<br />

Flow Termination: Pressure Support breaths are flow terminated when the flow decreases to a set<br />

percentage of the peak flow 24 delivered for that breath.<br />

Time Termination: Pressure Support breaths are time terminated 25 when the inspiratory time exceeds<br />

two breath periods, or when the inspiratory time exceeds the set Time Termination Limit 26 before the<br />

flow termination criteria is reached. The Pres. Support display will flash briefly after each time<br />

terminated breath.<br />

The Rise Time profile for Pressure Support breaths may be selected under Extended Features 26 .<br />

Pressure Support in NPPV Mode<br />

When NPPV mode is selected, the Pres. Support control is used to set the IPAP value.<br />

24 See Chapter 10 - Extended Features to set the Variable Flow Termination for Pressure Support breaths.<br />

25 Only Pressure Support breaths are flow and/or time terminated (Time Termination must be enabled).<br />

26 See Chapter 10 - Extended Features to set the Variable Time Termination or Flow Rise Time profile.<br />

Operator’s <strong>Manual</strong> <strong>LTV</strong> ® <strong>1200</strong> Ventilator Page 6-23<br />