WARNING<br />
Battery run time - When the battery reaches the BAT LOW level, the ventilator will only run for<br />
approximately 10 minutes before generating a battery empty alarm (BAT EMPTY). The approximate<br />
time shown is based on tests using the nominal settings, a new battery and a full 8 hour charge cycle<br />
as specified in Appendix A - Ventilator Specifications. Actual run time may be more or less depending<br />
on ventilator settings, patient demand, and battery age or condition. It is highly recommended that an<br />
alternate power source is connected PRIOR to the ventilator reaching the BAT EMPTY alarm<br />
condition to ensure continuous, uninterrupted patient ventilation<br />
INOP Alarm - If an INOP alarm occurs during operation, ventilate the patient using an alternative<br />
method, disconnect the ventilator, and immediately contact a certified Pulmonetic Systems’ service<br />
technician or Pulmonetic Systems.<br />
NO CAL Condition - Operation of the <strong>LTV</strong> ® <strong>1200</strong> ventilator under a NO CAL condition may result in<br />
inaccurate pressure and volume measurements. Should this condition occur, disconnect the patient<br />
from the ventilator, provide an alternative method of ventilation and immediately contact a certified<br />
Pulmonetic Systems’ service technician or Pulmonetic Systems.<br />
XDCR FAULT Alarm - Continued operation of the <strong>LTV</strong> ® <strong>1200</strong> ventilator with an activated XDCR<br />
FAULT alarm may result in inaccurate flow and volume measurements. Should this condition occur,<br />
disconnect the patient from the ventilator, provide an alternative method of ventilation and<br />
immediately contact a certified Pulmonetic Systems’ service technician or Pulmonetic Systems.<br />
Personal Injury and Electric Shock - Operation of the <strong>LTV</strong> ® <strong>1200</strong> ventilator if any of its panels have<br />
been removed may result in electrical shock to the patient or operator. All servicing must be<br />
performed by a certified Pulmonetic Systems’ service technician.<br />
Audible Alarms - Failure to immediately identify and correct audible alarm situations may result in<br />
serious patient injury.<br />
Equipment Malfunction or Failure - The <strong>LTV</strong> ® <strong>1200</strong> ventilator has alarms to notify operators of<br />
certain conditions and to cease operating upon detecting possible danger. In the event of equipment<br />
failure, all ventilator operators should have an alternative method of ventilation available and be fully<br />
familiar with emergency ventilation procedures.<br />
Improperly Functioning Ventilator - Operation of a ventilator that does not appear to be working<br />
properly may be hazardous. If the ventilator is damaged, fails Ventilator Checkout tests or<br />
malfunctions in any way, discontinue its use and immediately contact a certified Pulmonetic Systems’<br />
service technician or Pulmonetic Systems.<br />
Ventilator Checkout Tests – Be aware that gas is not delivered to the patient during these tests.<br />
Disconnect the patient from the ventilator and ventilate the patient using an alternative method before<br />
running the Ventilator Checkout tests.<br />
Ventilator Checkout and Maintenance Modes - The <strong>LTV</strong> ® <strong>1200</strong> ventilator does not deliver gas<br />
during the Ventilator Checkout mode (VENT CHECK) or Ventilator Maintenance mode (VENT<br />
MTNCE) and should not be used to ventilate a patient during these tests.<br />
Inspired Oxygen (FIO2) Concentration – If the patient has a variable respiratory rate, his/her minute<br />
ventilation will fluctuate. If exact concentrations of inspired oxygen (FIO2) are required to be delivered<br />
to the patient, it is recommended that an accurate oxygen analyzer with alarms be used.<br />
O2 Cylinder Duration Information - The accuracy of the displayed useable amount of oxygen<br />
remaining in an external O2 cylinder (O2 DUR hh:mm) is dependant on the precision of the pressure<br />
gauge used on the O2 cylinder and the accuracy of the information provided by the operator in the O2<br />
CYL DUR menu settings. The calculated/displayed useable amount of oxygen information is to be<br />
used for reference purposes only.<br />
Page 1-4 <strong>LTV</strong> ® <strong>1200</strong> Ventilator Operator’s <strong>Manual</strong><br />