WARNING<br />

Battery run time - When the battery reaches the BAT LOW level, the ventilator will only run for<br />

approximately 10 minutes before generating a battery empty alarm (BAT EMPTY). The approximate<br />

time shown is based on tests using the nominal settings, a new battery and a full 8 hour charge cycle<br />

as specified in Appendix A - Ventilator Specifications. Actual run time may be more or less depending<br />

on ventilator settings, patient demand, and battery age or condition. It is highly recommended that an<br />

alternate power source is connected PRIOR to the ventilator reaching the BAT EMPTY alarm<br />

condition to ensure continuous, uninterrupted patient ventilation<br />

INOP Alarm - If an INOP alarm occurs during operation, ventilate the patient using an alternative<br />

method, disconnect the ventilator, and immediately contact a certified Pulmonetic Systems’ service<br />

technician or Pulmonetic Systems.<br />

NO CAL Condition - Operation of the <strong>LTV</strong> ® <strong>1200</strong> ventilator under a NO CAL condition may result in<br />

inaccurate pressure and volume measurements. Should this condition occur, disconnect the patient<br />

from the ventilator, provide an alternative method of ventilation and immediately contact a certified<br />

Pulmonetic Systems’ service technician or Pulmonetic Systems.<br />

XDCR FAULT Alarm - Continued operation of the <strong>LTV</strong> ® <strong>1200</strong> ventilator with an activated XDCR<br />

FAULT alarm may result in inaccurate flow and volume measurements. Should this condition occur,<br />

disconnect the patient from the ventilator, provide an alternative method of ventilation and<br />

immediately contact a certified Pulmonetic Systems’ service technician or Pulmonetic Systems.<br />

Personal Injury and Electric Shock - Operation of the <strong>LTV</strong> ® <strong>1200</strong> ventilator if any of its panels have<br />

been removed may result in electrical shock to the patient or operator. All servicing must be<br />

performed by a certified Pulmonetic Systems’ service technician.<br />

Audible Alarms - Failure to immediately identify and correct audible alarm situations may result in<br />

serious patient injury.<br />

Equipment Malfunction or Failure - The <strong>LTV</strong> ® <strong>1200</strong> ventilator has alarms to notify operators of<br />

certain conditions and to cease operating upon detecting possible danger. In the event of equipment<br />

failure, all ventilator operators should have an alternative method of ventilation available and be fully<br />

familiar with emergency ventilation procedures.<br />

Improperly Functioning Ventilator - Operation of a ventilator that does not appear to be working<br />

properly may be hazardous. If the ventilator is damaged, fails Ventilator Checkout tests or<br />

malfunctions in any way, discontinue its use and immediately contact a certified Pulmonetic Systems’<br />

service technician or Pulmonetic Systems.<br />

Ventilator Checkout Tests – Be aware that gas is not delivered to the patient during these tests.<br />

Disconnect the patient from the ventilator and ventilate the patient using an alternative method before<br />

running the Ventilator Checkout tests.<br />

Ventilator Checkout and Maintenance Modes - The <strong>LTV</strong> ® <strong>1200</strong> ventilator does not deliver gas<br />

during the Ventilator Checkout mode (VENT CHECK) or Ventilator Maintenance mode (VENT<br />

MTNCE) and should not be used to ventilate a patient during these tests.<br />

Inspired Oxygen (FIO2) Concentration – If the patient has a variable respiratory rate, his/her minute<br />

ventilation will fluctuate. If exact concentrations of inspired oxygen (FIO2) are required to be delivered<br />

to the patient, it is recommended that an accurate oxygen analyzer with alarms be used.<br />

O2 Cylinder Duration Information - The accuracy of the displayed useable amount of oxygen<br />

remaining in an external O2 cylinder (O2 DUR hh:mm) is dependant on the precision of the pressure<br />

gauge used on the O2 cylinder and the accuracy of the information provided by the operator in the O2<br />

CYL DUR menu settings. The calculated/displayed useable amount of oxygen information is to be<br />

used for reference purposes only.<br />

Page 1-4 <strong>LTV</strong> ® <strong>1200</strong> Ventilator Operator’s <strong>Manual</strong><br />

p/n 18247-001, Rev. G