LOW O 2 PRES<br />

When the average oxygen inlet pressure is less than the minimum inlet pressure of 35 PSIG, the<br />

LOW O2 PRES alarm is generated. This alarm is only active when Low Pressure O2 Source is not<br />

selected and the oxygen concentration is set to greater than 21%.<br />

NOTE<br />

The <strong>LTV</strong> ® <strong>1200</strong> features an enhanced Low O2 Pressure alarm algorithm which allows for a brief,<br />

temporary drop of the O2 pressure supply while maintaining delivered O2 %.<br />

REMARQUE<br />

Le <strong>LTV</strong> est caractérisé par un algorithme amélioré ↔alarme de pression d’oxygène basse≈, lequel<br />

permet la chute brève et temporaire de l’approvisionnement de pression d’oxygène basse tout en<br />

maintenant le % d’oxygène administré.<br />

When a LOW O2 PRES alarm occurs:<br />

• The LOW O2 PRES message is flashed in the display window.<br />

• The O2% (O2 Flush) control display is flashed.<br />

• The audible alarm is sounded.<br />

To reset the LOW O2 PRES alarm:<br />

1) Push the Silence Reset button to silence the alarm.<br />

2) Reset the ventilator’s oxygen inlet pressure.<br />

3) Push the Silence Reset button to reset the alarm.<br />

Operator’s <strong>Manual</strong> <strong>LTV</strong> ® <strong>1200</strong> Ventilator Page 9-19<br />

p/n 18247-001, Rev. G