SBT OFF<br />

When the minutes preset in the SBT OP, MINUTES menu have elapsed and a Spontaneous<br />

Breathing Trial (SBT) mode of ventilation is terminated, the SBT OFF alarm is generated.<br />

For additional information concerning the Spontaneous Breathing Trial (SBT) mode of ventilation, see<br />

SBT (Spontaneous Breathing Trial) Operations in Chapter 10 - Extended Features.<br />

When a SBT OFF alarm is generated:<br />

• The SBT OFF message is flashed in the display window.<br />

• The audible alarm is sounded for 3 seconds.<br />

To clear the SBT OFF flashing display:<br />

1) Push the Silence Reset button.<br />

Operator’s <strong>Manual</strong> <strong>LTV</strong> ® <strong>1200</strong> Ventilator Page 9-31<br />

p/n 18247-001, Rev. G