Bi-directional Flow Transducer Differential Autozero<br />

Use this item to view the Bi-directional Flow Transducer Differential Autozero results and schedule<br />

Autozeros to be run.<br />

To view the Bi-directional Flow Transducer Differential Autozero results:<br />

1) The previous results, FDb xxxx P, are displayed. If the results are displayed as FDb xxxx -, the<br />

Bi-directional Flow Transducer is not installed on your unit. The final P indicates the previous zero<br />

results were within the required tolerance and the previous autozero passed. If a final F is<br />

displayed, the previous zero results were outside the required tolerance and the autozero failed.<br />

An asterisk indicates that an autozero is scheduled for the next breath.<br />

2) Turn the Set Value knob to display the EXIT option.<br />

3) Push the Select button.<br />

To schedule the Bi-directional Flow Transducer Differential Autozero:<br />

1) The previous results, FDb xxxx P, are displayed.<br />

2) Push the Select button. An asterisk appears, the pass / fail indicator is removed from the display<br />

and the autozero test is scheduled for the next breath.<br />

3) After the autozero is run on the next breath, the new autozero value and the pass / fail indicator<br />

are displayed.<br />

If the autozero fails, it will be automatically rescheduled for the next breath.<br />

Operator’s <strong>Manual</strong> <strong>LTV</strong> ® <strong>1200</strong> Ventilator Page 10-29<br />

p/n 18247-001, Rev. G