LOW PEEP<br />

When the patient circuit positive end expiratory pressure (PEEP) is less than the Low PEEP alarm<br />

setting, the LOW PEEP alarm is generated.<br />

• To prevent nuisance alarms, the LOW PEEP alarm is suspended for the first three (3) breaths or<br />

60 seconds (whichever is longer) after power on and ventilation begins 44 .<br />

When a LOW PEEP alarm occurs:<br />

• The LOW PEEP message is flashed in the display window.<br />

• The audible alarm is sounded.<br />

To reset the LOW PEEP alarm:<br />

1) Push the Silence Reset button once to silence the audible alarm.<br />

2) Push the Silence Reset button twice to reset the alarm (silences the audible alarm and clears the<br />

flashing display).<br />

• When the LOW PEEP alarm is reset (Silence Reset button pushed twice), the alarm is<br />

suspended for the next 60 seconds 44 .<br />

• The 60 second suspension of the LOW PEEP alarm is only enabled when the alarm is<br />

manually silenced/reset by pushing the Silence Reset button twice. The suspension is not<br />

enabled when the LOW PEEP alarm is automatically silenced/reset because the patient’s<br />

PEEP no longer exceeds the set LOW PEEP alarm value 44 .<br />

44 The Low PEEP alarm suspension is only available with software version 05.07 or higher installed.<br />

Page 9-20 <strong>LTV</strong> ® <strong>1200</strong> Ventilator Operator’s <strong>Manual</strong><br />

p/n 18247-001, Rev. G