MR Conditional System – Conditions for Operation<br />
The MR Conditional <strong>LTV</strong> ® <strong>1200</strong> System is suitable for use in both 1.5 and 3.0 Tesla (not to exceed<br />
3.0 Tesla static magnetic field) shielded magnetic scanners when operated under the following<br />
conditions:<br />
• The <strong>LTV</strong> ® <strong>1200</strong> MR Conditional System shall always be used as an assembled system, using<br />
all of the components (MR Conditional <strong>LTV</strong> ® <strong>1200</strong> ventilator, MR Conditional Floor Stand, MR<br />
Safe 15ft Patient Circuit, and the AC adapter).<br />
• Put the MR Conditional <strong>LTV</strong> ® <strong>1200</strong> System outside the 100 Gauss field line (at least 9ft /<br />
2.74m from the bore opening) and in a position where the operator can always see the <strong>LTV</strong> ®<br />
<strong>1200</strong>. Always lock the two (2) locking caster wheels on the ventilator floor stand when used in<br />
the MR environment.<br />
• Use only MR Conditional, non-magnetic (e.g., aluminum) oxygen cylinders, and MR<br />
Conditional oxygen regulators if O2 is required.<br />
• Follow all institutional guidelines, MR Conditional procedures and precautions, and the <strong>LTV</strong> ®<br />
<strong>1200</strong> Operator’s <strong>Manual</strong> when operating in the MR environment.<br />
WARNING<br />
• Only the MR Conditional <strong>LTV</strong> ® <strong>1200</strong> System (which includes the MR Conditional <strong>LTV</strong> ® <strong>1200</strong><br />
ventilator, MR Conditional Floor Stand, MR Safe 15ft Patient Circuit, and the AC adapter) may<br />
be used in the MR environment. To avoid patient harm and/or equipment damage, do not<br />
operate the MR Conditional <strong>LTV</strong> ® <strong>1200</strong> System in the MR environment if any system component<br />
is missing or not installed.<br />
• Failure to follow all warnings, the <strong>LTV</strong> ® <strong>1200</strong> Operator’s <strong>Manual</strong>, and all MR Conditional<br />
procedures and precautions when the MR Conditional <strong>LTV</strong> ® <strong>1200</strong> System is used may cause<br />
patient and/or operator harm and damage to the ventilator may occur. Contact Pulmonetic<br />
Systems with any questions.<br />
• Remove the <strong>LTV</strong> ® Carrying Strap (if attached) located at the top of the ventilator before entering<br />
the MR environment.<br />
AVERTISSEMENT<br />
• Dans un milieu de RM, seul le système MR Conditional <strong>LTV</strong> ® <strong>1200</strong> (qui inclut le ventilateur MR<br />
Conditional <strong>LTV</strong> ® <strong>1200</strong>, le support de plancher MR Conditional, le circuit de patient de 4,6 m (15<br />
pi) MR Safe et l'adaptateur de c.a.) peut être utilisé dans le milieu de RM. Pour éviter tout<br />
danger pour le patient et/ou tout dommage à l'équipement, ne faites pas fonctionner le système<br />
MR Conditional <strong>LTV</strong> ® <strong>1200</strong> dans un milieu de RM si tout composant du système manque ou<br />
n'est pas installé.<br />
• Ne pas suivre tous les avertissements, le manuel de l'opérateur <strong>LTV</strong> ® <strong>1200</strong> et toutes les<br />
précautions et procédures conditionnelles de RM lorsque le système MR Conditional <strong>LTV</strong> ® <strong>1200</strong><br />
est utilisé pourrait mener à des problèmes pour le patient et/ou pour l'opérateur et des<br />
dommages au ventilateur. Contactez Pulmonetic Systems pour toutes questions.<br />
• Retirez la sangle de transport <strong>LTV</strong> ® (si elle est installée) qui se trouve sur le dessus du<br />
ventilateur avant d'entrer dans le milieu de RM.<br />
Page 16-2 <strong>LTV</strong> ® <strong>1200</strong> Ventilator Operator’s <strong>Manual</strong><br />
p/n 18247-001, Rev. G