High Pressure Limit<br />

Use the High Pressure Limit to establish the maximum pressure permitted the patient circuit. When<br />

this limit is reached:<br />

• A HIGH PRES alarm is displayed<br />

• The audible alarm is sounded<br />

• Inspiration is terminated and exhalation begins<br />

The turbine is stopped to allow the circuit pressure to evacuate when the high pressure condition<br />

persists for more than four times the set inspiratory time or more than 3.0 seconds, whichever is less.<br />

To set the High Pressure Limit:<br />

1) Push the High Pres. Limit button.<br />

2) Change the setting using the Set Value knob.<br />

Range: 5 - 100 cmH2O<br />

Page 6-4 <strong>LTV</strong> ® <strong>1200</strong> Ventilator Operator’s <strong>Manual</strong><br />

p/n 18247-001, Rev. G