SBT > f/Vt<br />

When the Spontaneous Breathing Trial (SBT) mode of ventilation is on and the monitored f/Vt (Total<br />

Breath Rate divided by the Average Exhaled Tidal Volume) is greater than the HIGH f/Vt setting (high<br />

limit of the f/Vt range) for 30 seconds, the SBT > f/Vt alarm* is generated.<br />

For additional information concerning the Spontaneous Breathing Trial (SBT) mode of ventilation, see<br />

SBT (Spontaneous Breathing Trial) Operations in Chapter 10 - Extended Features.<br />

When a SBT > f/Vt alarm is generated:<br />

• The SBT > f/Vt message is flashed in the display window.<br />

• The audible alarm is sounded.<br />

To reset the SBT > f/Vt alarm:<br />

1) Push the Silence Reset button once to silence the audible alarm.<br />

2) Push the Silence Reset button twice to reset the alarm (silences the audible alarm and clears<br />

the flashing display).<br />

• When the SBT > f/Vt alarm is reset (Silence Reset button pushed twice), the alarm is<br />

suspended for the next 60 seconds.<br />

• The 60-second suspension of the SBT > f/Vt alarm is only enabled when the alarm is<br />

manually silenced/reset by pushing the Silence Reset button twice. It is not enabled<br />

when the SBT > f/Vt alarm is automatically silenced/reset because the patient’s breath<br />

rate is no longer greater than the set HIGH f/Vt alarm value.<br />

• When an SBT alarm has been active in excess of 5 minutes, the SBT mode of ventilation<br />

is terminated, the ventilator clears the alarm status, silences the audible alarm and<br />

returns to its previous mode of ventilation / settings. To remove the flashing SBT alarm<br />

message from the display window, push the Silence Reset button.<br />

Page 9-30 <strong>LTV</strong> ® <strong>1200</strong> Ventilator Operator’s <strong>Manual</strong><br />