Valve Home Position<br />

Use the Valve Home Position option to view the home position for the <strong>LTV</strong>’s flow valve. The home<br />

position is displayed as: VHome xxx where xxx is the home position for the valve installed in the<br />

ventilator.<br />

The home position is determined by the revision of the flow valve and is set when the ventilator is<br />

manufactured or when the flow valve is replaced by a certified Pulmonetic Systems’ service<br />

technician.<br />

Set Defaults<br />

The SET DEFAULTS option is only displayed and accessed through the Ventilator Checkout menu<br />

(VENT CHECK) or Ventilator Maintenance menu (VENT MTNCE) and is used to reset user settable<br />

Controls and Extended Features settings to their factory-set default values. See Chapter 9 - Ventilator<br />

Alarms, DEFAULTS for factory-set default values.<br />

To enable the Ventilator Checkout menu:<br />

To enable the Ventilator Checkout menu, the patient must be disconnected from the ventilator<br />

(ventilate the patient using an alternative method of ventilation), the ventilator must be turned off, and<br />

a special power on sequence used to turn it back on. See Chapter 11 - Ventilator Checkout Tests for<br />

important information and instructions, prior to proceeding.<br />

To set the default values:<br />

1) When the VENT CHECK menu is displayed, turn the Set Value knob until VENT OP is displayed<br />

and push Select.<br />

2) Turn the Set Value knob until DEFAULTS is displayed and push Select. SET DEFAULTS will<br />

be displayed.<br />

3) Push Select while SET DEFAULTS is displayed. DEFAULTS will be displayed.<br />

• Except for the Language selected and the Date/Time settings and format, all user settable<br />

Controls and Extended Features options are reset to their factory-set default values.<br />

• A DEFAULTS event is recorded in the Event Trace log (see Appendix E - Event Trace for<br />

additional information) along with the date and time the settings were reset.<br />

To exit the Ventilator Checkout menu and enter normal ventilation mode:<br />

1) Turn the Set Value knob through the Ventilator Operations sub-menus until EXIT is displayed,<br />

and push the Select or Control Lock button. VENT OP will be displayed.<br />

2) When VENT OP is displayed, turn the Set Value knob until EXIT is displayed, and push the<br />

Select or Control Lock button.<br />

• POST will be performed, the ventilator will begin ventilation using the factory set default<br />

settings and a DEFAULTS SET alarm will be generated (see Chapter 9 - Ventilator Alarms,<br />

DEFAULTS SET for additional information and instructions to reset the DEFAULTS SET<br />

alarm).<br />

Page 10-20 <strong>LTV</strong> ® <strong>1200</strong> Ventilator Operator’s <strong>Manual</strong><br />

p/n 18247-001, Rev. G