Cautions<br />

CAUTION<br />

Ventilator Sterilization – To avoid irreparable damage to the <strong>LTV</strong> ® <strong>1200</strong> ventilator, do not attempt to<br />

sterilize it.<br />

Cleaning Agents – To avoid damaging the ventilator’s plastic components and front panel, do not<br />

use cleaning agents containing ammonium chloride, other chloride compounds, more than 2%<br />

glutaraldehyde, phenols, or abrasive cleaners.<br />

Ventilator Immersion - Do not immerse the ventilator in liquids.<br />

Reusable Patient Circuit Components - To avoid degradation of the reusable patient circuit<br />

components, do not exceed the following constraints:<br />

• 50 cleaning cycles or 1 year (whichever comes first)<br />

Steam Autoclave:<br />

• Pressure: 20 PSIG<br />

• Temperature: 275°F (135°C)<br />

• Time: 6 minutes<br />

Liquid Sterilizing Agent:<br />

Do not use any of the following solutions to clean, disinfect, or sterilize the patient circuit:<br />

• Ketone • Chlorinated solutions<br />

• Phenol (>5%) • Chlorinated hydrocarbons<br />

• Inorganic acids • Aromatic hydrocarbons<br />

• Formaldehyde • Hypochlorite<br />

• Liquid agents containing more than<br />

2% glutaraldehyde<br />

Pasteurization:<br />

• A 30-minute warm water detergent and a 30-minute 165°F (74°C) hot water cycle.<br />

• Drying in a sterile drier for more than 1 hour or 140°F (59°C).<br />

Gas (ETO):<br />

• Temperature: 131°F (55°C)<br />

Differential Pressure Ports - A low pressure air nozzle with flow less than 10 liters per minute<br />

should be used for cleaning the differential pressure ports.<br />

Exhalation Valve Cleaning - Do not pour or spray liquid cleaners into the exhalation valve.<br />

Patient Wye Installation – After cleaning, install the patient wye in the patient circuit so the proximal<br />

sense lines are oriented up while operating.<br />

Page 1-10 <strong>LTV</strong> ® <strong>1200</strong> Ventilator Operator’s <strong>Manual</strong><br />

p/n 18247-001, Rev. G