Inspiratory Hold<br />

An Inspiratory Hold maneuver holds the inspiratory phase of a delivered breath for a duration<br />

sufficient to determine Δ Pres pressure and static lung compliance of the patient.<br />

To perform the Inspiratory Hold maneuver:<br />

1) Push the Insp/Exp (Inspiratory/Expiratory) Hold control button once and the display window will<br />

toggle from normal monitor display to INSP HOLD.<br />

2) Push and hold the Insp/Exp (Inspiratory/Expiratory) Hold button during a volume inspiration.<br />

• The ventilator will perform an Inspiratory Hold on the next Volume breath.<br />

• P Plat 14 --- will be displayed in the display window.<br />

• All buttons that are not lockable will operate normally.<br />

• All buttons that are lockable will be ignored.<br />

3) Continue holding the button until the Volume inspiration is completed. During the maneuver:<br />

• The exhalation valve will remain closed.<br />

• Flow will be set to 0 LPM.<br />

• P Plat xxx will be displayed in the display window, where xxx is the real time circuit<br />

pressure.<br />

• The breath period will remain in inspiration phase so no breath triggers are allowed.<br />

• DISC/SENSE and HIGH PRES alarms will terminate the maneuver.<br />

4) Release the button when the pressure setting is P Plat (or when 6.0 seconds elapse, whichever<br />

comes first):<br />

• The exhalation valve will be opened and a normal exhalation phase will begin.<br />

• The display will cycle every 2 seconds between Δ Pres xxx where xxx is the change in<br />

pressure 15 , C Static xxx where xxx is the static compliance 16 and<br />

P Plat xxx where xxx is the plateau pressure.<br />

NOTE<br />

Breath period timing and apnea timing will be suspended while the maneuver is performed. As a<br />

result, the apnea alarm will not alarm during the maneuver.<br />

REMARQUE<br />

La synchronisation de la période de respiration et la synchronisation de l’apnée sont interrompues<br />

pendant la manœuvre. Ainsi, l’alarme d’apnée ne se déclenche pas au cours de la manœuvre.<br />

Range: P Plat 0 - 100 cmH2O<br />

Δ Pres 0 - 100 cmH2O<br />

C Static 1 – 999 ml/cmH2O<br />

14 “P Plat” is Plateau pressure reached during Inspiratory Hold maneuver.<br />

15 Δ Pres is calculated as P Plat Pressure – PEEP measured from previous breath.<br />

16 C Static is calculated as Set Delivered Volume / Δ Pres.<br />