To enable the Ventilator Checkout menu:<br />

1) Disconnect the patient from the ventilator and ventilate the patient using an alternative method.<br />

2) Turn the ventilator off.<br />

3) Ensure that the AC Adapter is connected to a valid AC power source and verify that the External<br />

Power and Charge Status LEDs are illuminated.<br />

4) Push and hold the Select button. Continue to hold the Select button and press the On/Standby<br />

button. REMOVE PTNT should be displayed; if not, power down the ventilator and repeat steps<br />

2 through 4.<br />

5) An audible alarm (alternating on/off tone) will sound while REMOVE PTNT is displayed.<br />

6) Clear the alarm by pressing the Silence Reset button. The audible alarm silences, and the<br />

display changes to VENT CHECK.<br />

7) Push the Select button. The first Ventilator Checkout Test, ALARM, is displayed.<br />

Page 11-2 <strong>LTV</strong> ® <strong>1200</strong> Ventilator Operator’s <strong>Manual</strong><br />

p/n 18247-001, Rev. G