Low Minute Volume<br />

The Low Minute Volume alarm sets the minimum expected exhaled Minute Volume. The exhaled<br />

Minute Volume is recalculated after every breath. If the Minute Volume does not meet or exceed the<br />

Low Minute Volume setting:<br />

• A LOW MIN VOL alarm is displayed<br />

• The audible alarm is sounded<br />

To set the Low Minute Volume alarm:<br />

1) Push the Low Min. Vol. button.<br />

2) Change the setting using the Set Value knob.<br />

Range: Off, 0.1 - 99 L<br />

WARNING<br />

Low Minute Volume Control Settings - The Low Min. Vol. control should be set to its highest<br />

clinically appropriate value. If there is a clinical need to set the Low Minute Volume alarm to lower<br />

values or off (“- - -“), perform a clinical assessment to determine if an alternative monitor (i.e. a Pulse<br />

Oxymeter with an audible alarm, or a Cardio Respiratory Monitor) should be used.<br />

AVERTISSEMENT<br />

Réglages du contrôle de volume bas par minute - Le contrôle du volume bas par minute doit être<br />

ajusté à la plus haute valeur clinique appropriée. Si l’alarme de volume bas par minute doit être<br />

ajustée à des valeurs inférieures ou mise à l’arrêt (“- - -“) pour satisfaire aux besoins cliniques,<br />

effectuer une évaluation clinique afin de déterminer si l’utilisation d’un autre moniteur (c.-à-d.,<br />

sphygmo-oxymètre muni d’une alarme sonore ou un moniteur cardio-respiratoire) s’avère pertinente.<br />

Operator’s <strong>Manual</strong> <strong>LTV</strong> ® <strong>1200</strong> Ventilator Page 6-11<br />

p/n 18247-001, Rev. G