Symptoms Possible Causes What to Do<br />

Remote Alarm System PNT ASSIST option set to Set PNT ASSIST option to PULSE. See<br />

(dual tone system) NORMAL.<br />

page 10-6 for instructions.<br />

only generates one<br />

continuous tone.<br />

Defective Remote Alarm<br />

System.<br />

Contact Remote Alarm System<br />

manufacturer or service personnel.<br />

Internal problem with the Immediately contact a certified Pulmonetic<br />

ventilator.<br />

Systems’ service technician.<br />

Patient Assist Call Incorrect Patient Assist Establish whether the Patient Assist Call<br />

System does not work cable installed (Normally System is a Normally Open or Normally<br />

with the ventilator. Open versus Normally Closed system and verify the appropriate<br />

Closed system/cable Patient Assist Cable (Normally Open or<br />

mismatch)<br />

Normally Closed) is installed. See page C-<br />

14 for instructions.<br />

Defective or improper Check the Patient Assist Cable connection<br />

connections.<br />

between the ventilator’s Patient Assist Port<br />

and the Patient Assist Call System. See<br />

page C-14 for instructions.<br />

Defective Patient Assist<br />

cable.<br />

Replace Patient Assist Cable.<br />

Defective Patient Assist Contact Patient Assist Call System<br />

Call System.<br />

manufacturer or service personnel.<br />

Internal problem with the Immediately contact a certified Pulmonetic<br />

ventilator.<br />

Systems’ service technician.<br />

Patient Assist Call PNT ASSIST option set to Set PNT ASSIST option to NORMAL. See<br />

System generates a PULSE.<br />

page 10-6 for instructions.<br />

pulsating tone or light<br />

and manufacturers<br />

instructions indicate it<br />

should be a<br />

continuous tone or<br />

light.<br />

Defective Patient Assist<br />

Call System.<br />

Internal problem with the<br />

ventilator.<br />

Contact Patient Assist Call System<br />

manufacturer or service personnel.<br />

Immediately contact a certified Pulmonetic<br />

Systems’ service technician.<br />

Repeated SBT > f Total Breath Rate (f) Check SBT > f alarm value. See page 10alarms.<br />

exceeds the set SBT > f<br />

alarm value.<br />

27 for instructions.<br />

Repeated SBT < f<br />

alarms.<br />

Patient Circuit leak, causing<br />

autocycling.<br />

Internal problem with the<br />

ventilator.<br />

Total Breath Rate (f) is less<br />

than the set SBT < f alarm<br />

value.<br />

Do a Leak test and reseat or replace the<br />

leaking parts or connections. See page<br />

11-8 for instructions.<br />

If problem reoccurs, immediately contact a<br />

certified Pulmonetic Systems service<br />

technician.<br />

Check SBT < f alarm value. See page 10-<br />

27 for instructions.<br />

Patient Circuit leak. Do a Leak test and reseat or replace the<br />

leaking parts or connections. See page<br />

11-8 for instructions.<br />

Internal problem with the<br />

ventilator.<br />

If problem reoccurs, immediately contact a<br />

certified Pulmonetic Systems service<br />

technician.<br />

Operator’s <strong>Manual</strong> <strong>LTV</strong> ® <strong>1200</strong> Ventilator Page 15-23<br />

p/n 18247-001, Rev. G