Alarm Test<br />

Use the Alarm Test to verify that the audible alarm is working correctly.<br />

To run the Alarm Test:<br />

1) Push the Select button while ALARM is displayed.<br />

2) Verify that the audible alarm is sounded.<br />

• If a Patient Assist Call System or Remote Alarm is connected via the ventilator’s Patient<br />

Assist Port, verify the device also activates (audible/visual), as specified by its manufacturer.<br />

3) When the alarm has sounded for at least 2 seconds, push the Select button again.<br />

• The audible alarm is silenced and the next menu item is displayed.<br />

4) Verify a confirming audible chirp occurs after the alarm is silenced.<br />

If the Alarm Test fails, see Chapter 15 - Troubleshooting for more information.<br />

Operator’s <strong>Manual</strong> <strong>LTV</strong> ® <strong>1200</strong> Ventilator Page 11-3<br />

p/n 18247-001, Rev. G