O 2 Conserve<br />

Use the O2 Conserve option to minimize the use of oxygen from an O2 source while maintaining the<br />

FIO2 during inspiration.<br />

When CONSERVE ON is selected, the ventilator provides a bias flow of 0 lpm and automatically<br />

selects pressure triggering. Oxygen is supplied during inspiration, but not during expiration. O2<br />

Conserve ON reduces trigger sensitivity and should be used when the conservation of oxygen is<br />

necessary.<br />

When CONSERVE OFF is selected, the ventilator provides a bias flow of 10 lpm and automatically<br />

selects flow triggering. O2 Conserve OFF provides the best trigger sensitivity and should be used<br />

when the conservation of oxygen is not a priority.<br />

• The O2 Conserve option is not affected by turning the ventilator on or off. The ventilator<br />

retains the desired setting through power up and power down cycles.<br />

• If O2 Conserve is set to ON, the message BIAS FLO OFF will scroll through the<br />

alphanumeric display when the ventilator is in operation.<br />

To modify the O2 CONSERVE setting:<br />

1) Push the Select button while O2 CONSERVE is displayed.<br />

2) CONSERVE ON or CONSERVE OFF is displayed.<br />

3) Turn the Set Value knob until the desired state is displayed.<br />

4) Push the Select button.<br />

Options: ON or OFF<br />

WARNING<br />

O2 Conserve – When CONSERVE ON is selected:<br />

• The <strong>LTV</strong> ® <strong>1200</strong> automatically sets the bias flow to 0 lpm and selects pressure triggering. Certain<br />

patients may experience difficulty triggering a breath with a bias flow of 0 lpm.<br />

• Delivered FIO2 can be lower than the set value when tidal volumes are less than 100 ml.<br />

AVERTISSEMENT<br />

O2 Conserver – Quand “CONSERVE ON” est sélectionné :<br />

• Le <strong>LTV</strong> <strong>1200</strong> règle automatiquement le flux à 0 lpm et sélectionne le déclencheur de pression.<br />

Certains patients peuvent éprouver des la difficulté à déclencher une respiration avec un flux de 0<br />

l/min.<br />

• Le FI02 peut diminuer la valeur sélectionnée quand le volume courant est de moins de 100 ml.<br />

Operator’s <strong>Manual</strong> <strong>LTV</strong> ® <strong>1200</strong> Ventilator Page 10-13<br />

p/n 18247-001, Rev. G