Inspiratory / Expiratory Hold<br />

Pushing the Insp/Exp (Inspiratory/Expiratory) Hold control button causes the ventilator to toggle<br />

between the following messages in the display window. Each push causes the next item in sequence<br />

to be displayed:<br />

INSP HOLD<br />

EXP HOLD<br />

Normal monitor display<br />

While INSP HOLD or EXP HOLD is displayed:<br />

• The Insp/Exp Hold control button LED will flash on and off.<br />

• If the Insp/Exp Hold control button is not pushed within 60 seconds, the message will be<br />

removed and the LED will turn off.<br />

• Pushing the Select, Silence Reset or Control button will return the display to normal and the<br />

LED will stop flashing.<br />

Operator’s <strong>Manual</strong> <strong>LTV</strong> ® <strong>1200</strong> Ventilator Page 6-5<br />

p/n 18247-001, Rev. G