Symptoms Possible Causes What to Do<br />

Repeated SBT > f/Vt<br />

alarms.<br />

Repeated SBT < f/Vt<br />

alarms.<br />

Total Breath Rate (f)<br />

exceeds the set SBT > f/Vt<br />

alarm value.<br />

Check SBT > f/Vt alarm value. See page<br />

10-27 for instructions.<br />

Patient Circuit leak. Do a Leak test and reseat or replace the<br />

leaking parts or connections. See page<br />

11-8 for instructions.<br />

Internal problem with the<br />

ventilator.<br />

Total Breath Rate (f) is less<br />

than the set SBT < f/Vt<br />

alarm value.<br />

If problem reoccurs, immediately contact a<br />

certified Pulmonetic Systems service<br />

technician.<br />

Check SBT < f/Vt alarm value. See page<br />

10-27 for instructions.<br />

Patient Circuit leak. Do a Leak test and reseat or replace the<br />

leaking parts or connections. See page<br />

11-8 for instructions.<br />

Internal problem with the<br />

ventilator.<br />

Checkout Test Failures<br />

Symptoms Possible Causes What to Do<br />

Alarm Test<br />

Audible alarm level<br />

excessive.<br />

Alarm Test<br />

Audible alarm too soft.<br />

Alarm Test<br />

Alarm does not sound.<br />

Alarm Test<br />

Confirming audible<br />

chirp does not sound.<br />

If problem reoccurs, immediately contact a<br />

certified Pulmonetic Systems service<br />

technician.<br />

Alarm volume set too high. Set the alarm volume under the Extended<br />

Features Menu. (See page 10-3 for an<br />

explanation of the ALARM VOL feature.)<br />

Alarm volume set too low. Set the alarm volume under the Extended<br />

Features Menu. (See page 10-3 for an<br />

explanation of the ALARM VOL feature.)<br />

Alarm sounder blocked. Check the alarm sounder opening in the<br />

right side of the ventilator to verify the<br />

opening is not blocked.<br />

Internal problem with the<br />

ventilator.<br />

Immediately contact a certified Pulmonetic<br />

Systems’ service technician.<br />

Alarm sounder blocked. Check the alarm sounder opening in the<br />

right side of the ventilator to verify the<br />

opening is not blocked.<br />

Internal problem with the<br />

ventilator.<br />

Audible alarm did not sound<br />

long enough before test<br />

was terminated.<br />

Immediately contact a certified Pulmonetic<br />

Systems’ service technician.<br />

Repeat the Alarm Test and allow audible<br />

alarm to sound for at least 2 seconds<br />

before pushing the Select button. (See<br />

Alarm Test in Chapter 11 for instructions.)<br />

Page 15-24 <strong>LTV</strong> ® <strong>1200</strong> Ventilator Operator’s <strong>Manual</strong><br />