Pressure Control Breaths<br />

For Pressure Control breaths 9 , flow is delivered to elevate the circuit pressure to the Pressure Control<br />

setting and maintain it at that pressure for the set Inspiratory Time. Pressure Control breaths may be<br />

machine or assist type breaths.<br />

Adjusting the Rise Time Profile changes the flow and pressure waveforms for Pressure Control<br />

breaths.<br />

9 Pressure Control and Pressure Support breaths do compensate for PEEP. For example, a Pressure Control<br />

setting of 20cmH2O and a PEEP setting of 10cmH2O results in a Peak Inspiratory Pressure (PIP) of 30cmH2O<br />

(20 cmH20 over the set PEEP).<br />

Operator’s <strong>Manual</strong> <strong>LTV</strong> ® <strong>1200</strong> Ventilator Page 3-3<br />

p/n 18247-001, Rev. G