Pressure Control<br />

This control establishes the target pressure above the PEEP setting for Pressure Control breaths.<br />

NOTE<br />

Pressure Control and Pressure Support breaths do compensate for PEEP. Delivered pressure is<br />

controlled by the Pressure Control setting and is affected by the PEEP setting. For example, a<br />

Pressure Control setting of 20cmH2O and a PEEP setting of 10cmH2O results in a Peak Inspiratory<br />

Pressure (PIP) of 30cmH2O.<br />

REMARQUE<br />

Le contrôle de la pression et la pression de support des respirations compensent la PEP. La pression<br />

administrée est contrôlée par le réglage du contrôle de la pression et elle dépend du réglage PEP.<br />

Par exemple, un réglage du contrôle de la pression de 20 cmH2O et un réglage PEP de 10 cmH2O<br />

donnent une pression inspiratoire maximale (PImax) de 30 cmH2O.<br />

The inspiratory time for the Pressure Control breath is determined by the Inspiratory Time setting.<br />

The ventilator controls inspiratory flow to maintain the set circuit pressure for the set time.<br />

To set the Pressure Control level:<br />

1) Push the Pres. Control button.<br />

2) Change the setting using the Set Value knob.<br />

To select Pressure Control:<br />

1) Toggle the Volume Pressure mode to select Pressure ventilation 21 .<br />

Range: 1 - 99 cmH2O<br />

Flow Termination for Pressure Control breaths may be enabled under Extended Features 22 . If flow<br />

termination is enabled, the Pres. Control display will flash briefly after each flow terminated breath.<br />

The Rise Time profile for Pressure Control breaths may be selected under Extended Features 23 .<br />

21 See Chapter 6 - Controls for more information on how to select Pressure ventilation.<br />

22 See Chapter 10 - Extended Features for how to set the Variable Flow Termination percentage and enable<br />

Flow Termination for Pressure Control breaths.<br />

23 See Chapter 10 - Extended Features for how to set the Flow Rise Time profile.<br />

Page 6-22 <strong>LTV</strong> ® <strong>1200</strong> Ventilator Operator’s <strong>Manual</strong><br />

p/n 18247-001, Rev. G