Control Range Tolerance<br />

PIP LED Display On, Off n/a<br />

Pressure Control 1 to 99 cmH2O ± 2 cmH2O or 8% whichever is<br />

greater, steady-state only<br />

Pressure Control Flow<br />

Termination<br />

On, Off n/a<br />

Pressure Support “--”, 1 to 60 cmH2O ± 2 cmH2O or 8% whichever is<br />

greater, steady-state only.<br />

Set Date 01/01/1998 to 12/31/2097 n/a<br />

Set Time 00:00:00 to 23:59:59 n/a<br />

Sensitivity 1 to 9 Lpm, “-” + 1/- 0.5 lpm for setting of 1; ± 1<br />

lpm for all other settings.<br />

Tidal Volume 50 to 2000 ml ± 10% or 10 ml, whichever is<br />

greater for temperatures from 20°C<br />

to 30°C only, standard atmospheric<br />

pressure<br />

Variable Flow<br />

Termination<br />

10% to 40% ± 15% or 2 lpm whichever is greater<br />

Variable Rise Time 1 to 9 0.1 to 1.0 sec<br />

Variable Time<br />

Termination<br />

0.3 to 3.0 sec ± 0.1 sec<br />

Bias Flow 0 lpm or 10 lpm during<br />

exhalation 64<br />

SBT Start On - Off n/a<br />

± 10% or 1 lpm, whichever is<br />

greater<br />

SBT PS 0-30 cmH2O ± 2 cmH2O or 8% whichever is<br />

greater, steady-state only.<br />

SBT PEEP 0-20 cmH2O ± 2 cmH2O or 10%, whichever is<br />

greater<br />

SBT FIO2 21-100% O2 % mean: 21% to 50%: ± 3<br />

51% to 95%: ± 5<br />

keep steady-state only<br />

SBT O2 minutes 15-120 minutes n/a<br />

SBT high f/Vt 70-120 ± 20%<br />

SBT low f/Vt 5-90 ± 20%<br />

SBT High f 15-80 BPM ± 1 bpm or within 5%, whichever is<br />

greater<br />

SBT Low f 0-40 BPM ± 1 bpm or within 5%, whichever is<br />

greater<br />

Display f/Vt On - Off n/a<br />

64 0 lpm when O2 Conserve ON is selected or 10 lpm when O2 Conserve OFF is selected.<br />