Storage and Operating Conditions (cont.)<br />

Specification<br />

Oxygen Inlet<br />

Tolerance<br />

DISS or NIST Connector Inlet Pressure Range 40 to 80 PSIG ± 2 PSIG<br />

Tapered Tubing Connector Inlet Pressure Range 0 to 10 PSIG ± 2 PSIG<br />

Shock and Vibration<br />

The ventilator is designed to withstand shock and vibration in accordance with relevant<br />

requirements set forth in the following standards:<br />

IEC 68-2-27 Shock<br />

IEC 68-2-6 Vibration<br />

IEC 68-2-34 Vibration<br />

MIL-STD-810E Shock, Ground Transport and Helicopter Transport Vibration<br />

Spillage<br />

The ventilator resists fluid spillage when tested in accordance with the relevant standards<br />

specified in IEC 601-1 Clause 44.3.<br />

External Surface Temperature<br />

External surfaces < 50°C, ambient temperature of 35°C n/a<br />

Communications<br />

Port Connector Specification<br />

Communications RS232, DB9 connector Protocol Options: Data, Monitor, Printer,<br />

Modem<br />

Patient Assist Call /<br />

Remote Alarm<br />

Equipment Classification<br />

RJ11-4 Closed contact resistance: ≤ 1 ohm<br />

Classification The ventilator is rated as Class II equipment per IEC 601-1 Clause 6.11<br />

Type The ventilator is specified as Type BF equipment per IEC 601-1 Clause 6.11<br />

Operator’s <strong>Manual</strong> <strong>LTV</strong> ® <strong>1200</strong> Ventilator Page A-7<br />

p/n 18247-001, Rev. G