Cleaning or replacing the Inlet Filter<br />

To clean the Inlet Filter:<br />

1) Remove the Inlet Filter by squeezing the foam filter gently with your fingers and pulling it out.<br />

2) Gently bathe the filter in a solution of mild detergent and warm water.<br />

3) Rinse thoroughly in warm water.<br />

4) Examine the filter for excessive wear or damage. Discard and replace with a new filter if<br />

necessary.<br />

5) Allow the filter to thoroughly air dry before reinstallation.<br />

6) Reinstall the filter.<br />

CAUTION<br />

Wet or Damp Filters - Do not install a wet or damp filter into the <strong>LTV</strong> ® <strong>1200</strong> ventilator. This could<br />

damage the ventilator.<br />

ATTENTION<br />

Filtres mouillés ou humides - Ne pas installer des filtres mouillés ou humides dans les ventilateurs<br />

de la <strong>1200</strong> <strong>LTV</strong> ® . Cela pourrait endommager le ventilateur.<br />

Operator’s <strong>Manual</strong> <strong>LTV</strong> ® <strong>1200</strong> Ventilator Page 13-3<br />