Symptoms Possible Causes What to Do<br />

continued…<br />

Monitored volume is<br />

low.<br />

Delivered volume is<br />

low.<br />

Delivered volume is<br />

half the set volume.<br />

Delivered pressure is<br />

low, PEEP is low,<br />

ventilator is<br />

autocycling.<br />

Delivered pressure is<br />

low.<br />

Monitored pressure is<br />

low.<br />

Exhalation drive line<br />

leaking or loose.<br />

Exhalation valve<br />

leaking during<br />

inspiration.<br />

Sense lines are<br />

reversed.<br />

Leak Compensation is<br />

not on.<br />

Check the exhalation drive line at both the<br />

ventilator and exhalation valve ends. Verify the<br />

line is securely seated and not leaking.<br />

Verify the exhalation valve is not leaking during<br />

inspiration. If it is leaking, open the exhalation<br />

valve and remove the diaphragm. Reseat the<br />

diaphragm. See page 13-10 for a diagram of<br />

correct exhalation valve assembly.<br />

If necessary, replace the exhalation diaphragm,<br />

or exhalation valve with a new one.<br />

The sense lines are not designed to be removed<br />

from either the wye or the luer fittings. If the<br />

sense lines have been removed and replaced<br />

incorrectly, they may not seal correctly when<br />

replaced. Replace the patient wye and sense<br />

lines with a known good assembly.<br />

Verify that the Leak Compensation extended<br />

features option is set to On (default setting is on).<br />

See page 10-12 for instructions.<br />

Failed autozero. Perform an autozero under XDCR ZERO. See<br />

page 10-28 for more information.<br />

Failed calibration or<br />

internal problem with<br />

the ventilator.<br />

VHome setting does<br />

not match flow valve.<br />

Immediately contact a certified Pulmonetic<br />

Systems’ service technician.<br />

Immediately contact a certified Pulmonetic<br />

Systems’ service technician.<br />

Circuit leak. Run a leak test and reseat or replace the leaking<br />

parts or connections. See page 11-8 for<br />

instructions.<br />

Page 15-8 <strong>LTV</strong> ® <strong>1200</strong> Ventilator Operator’s <strong>Manual</strong><br />

p/n 18247-001, Rev. G