O 2 Flush<br />

Use the O2 Flush option to elevate the delivered FIO2 to 100% for a preset period of time.<br />

To initiate an O2 Flush:<br />

1) Push and hold the O2 % (O2 Flush) button (FIO2) on the ventilator front panel for three seconds to<br />

initiate the elevation (ramp up) of delivered FIO2 to 100% for the preset number of minutes.<br />

• The FIO2 percentile displayed will change to 100 and the O2 Flush maneuver will start<br />

immediately (regardless of the current ventilation mode, breath rate or phase).<br />

• O2 Flush will not be initiated if Low Pressure O2 Source is selected.<br />

• O2 Flush will stop when the preset minutes have elapsed or the O2 % (O2 Flush) button is<br />

pushed again.<br />

• When stopped, the delivered FIO2 percentile will return (ramp down) to the preset O2 %<br />

setting.<br />

To modify the O2 Flush setting:<br />

1) Push the Select button while O2 FLUSH is displayed and xxx min is displayed.<br />

2) Turn the Set Value knob until the desired amount of minutes is displayed and push the Select<br />

button. The O2 Flush time period is set.<br />

Range: 1 to 3 minutes, in increments of 1<br />

Page 10-14 <strong>LTV</strong> ® <strong>1200</strong> Ventilator Operator’s <strong>Manual</strong><br />