Mechanical Controls<br />

Control Range Tolerance<br />

Over Pressure Relief ≤125 cmH2O N/A<br />

Sub-Ambient Relief Pressure Drop: ≤ 5 cmH2O at 50 lpm<br />

Internal Compliance<br />

Monitors<br />

Compliance < 0.1 mL/cm<br />

Monitor Range Tolerance<br />

Calculated Peak Flow 10 to 100 lpm 2 lpm or ± 10%, whichever is greater<br />

Exhaled Tidal Volume 0 to 4000 ml ± 15% or 15 ml, whichever is greater<br />

I:E Ratio, Measured 99:1 and 1:99<br />

Based on the measured<br />

inspiratory / exhalation times<br />

I:E Ratio, Calculated 1:99 to 4.0:1 based on set<br />

breath rate and inspiratory<br />

time<br />

Accuracy for times are ± 50 ms or 5%,<br />

whichever is greater<br />

± 5%<br />

Mean Airway Pressure 0 to 99 cmH2O ± 2 cmH2O or 10%, whichever is greater<br />

O2 Cylinder Duration 0 - 99 hours and 59 minutes - 0 / + 40%<br />

Peak Inspiratory<br />

Pressure<br />

0 to 120 cmH2O ± 2 cmH2O or 5%, whichever is greater<br />

PEEP 0 to 99 cmH2O ± 2 cmH2O or 10%, whichever is greater<br />

Total Breath Rate 0 to 250 breaths per minute ± 1 bpm or within 5% of the breath<br />

period, whichever is greater<br />

Total Minute Volume 0 to 99.9 liters ± 15%, or the measured total breath rate<br />

times 15 ml, whichever is greater<br />

SBT Minutes 15 – 120 minutes ± 0.1 seconds<br />

f/Vt<br />

and<br />

f<br />

Button Controls<br />

0 – 4000<br />

0 – 250bpm<br />

± 20%<br />

Total breath rate: ± 1 bpm or within 5%<br />

of breath period, whichever is greater<br />

Display Function<br />

Control Lock Locks front panel controls, can be set to Easy or Hard unlocking<br />

<strong>Manual</strong> Breath Generates a machine breath<br />

Standby/On Puts ventilator in On or Standby state<br />

Low Pressure O2 Source Selects Low Pressure O2 Source<br />

Silence Reset Silences and resets alarms<br />

Operator’s <strong>Manual</strong> <strong>LTV</strong> ® <strong>1200</strong> Ventilator Page A-5<br />

p/n 18247-001, Rev. G

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!