LOW MIN VOL<br />

When the exhaled minute volume (VE) is less than the Low Minute Volume setting, the LOW MIN<br />

VOL alarm is generated.<br />

• To prevent nuisance alarms, the LOW MIN VOL alarm is suspended for the first 20 seconds of<br />

ventilator operation after power up and passing the Power On Self Tests.<br />

When a LOW MIN VOL alarm occurs:<br />

• The LOW MIN VOL message is flashed in the display window.<br />

• The Low Min. Vol. control display is flashed.<br />

• The audible alarm is sounded.<br />

To reset the LOW MIN VOL alarm:<br />

1) Push the Silence Reset button twice.<br />

WARNING<br />

Low Minute Volume Control Settings - The Low Min. Vol. control should be set to its highest<br />

clinically appropriate value. If there is a clinical need to set the Low Minute Volume alarm to lower<br />

values or off (“- - -“), perform a clinical assessment to determine if an alternative monitor (i.e. a Pulse<br />

Oxymeter with an audible alarm, or a Cardio Respiratory Monitor) should be used.<br />

AVERTISSEMENT<br />

Réglages du contrôle de volume bas par minute - Le contrôle du volume bas par minute doit être<br />

ajusté à la plus haute valeur clinique appropriée. Si l’alarme de volume bas par minute doit être<br />

ajustée à des valeurs inférieures ou mise à l’arrêt (“- - -“) pour satisfaire aux besoins cliniques,<br />

effectuer une évaluation clinique afin de déterminer si l’utilisation d’un autre moniteur (c.-à-d.,<br />

sphygmo-oxymètre muni d’une alarme sonore ou un moniteur cardio-respiratoire) s’avère pertinente.<br />

Page 9-18 <strong>LTV</strong> ® <strong>1200</strong> Ventilator Operator’s <strong>Manual</strong><br />

p/n 18247-001, Rev. G