Patient Breathing Circuit – Connection Instructions<br />

1) Connect the main breathing tube to the 22 mm outlet port on the right side of the ventilator.<br />

2) Connect the two exhalation flow transducer sense lines to the ports marked Flow Xducer on the<br />

right side of the ventilator. These are non-interchangeable Luer fittings.<br />

3) Connect the exhalation valve drive line to the port marked Exh Valve on the right side of the<br />

ventilator.<br />

CAUTION<br />

Patient Wye Installation – After cleaning, install the patient wye in the patient circuit so the proximal<br />

sense lines are oriented up while operating.<br />

ATTENTION<br />

Installation de la soupape d'expiration - Après le nettoyage, installez la soupape d'expiration dans<br />

le circuit du patient de sorte que les lignes de détection soient alignées vers le haut pendant<br />

l’opération.<br />

Page C-10 <strong>LTV</strong> ® <strong>1200</strong> Ventilator Operator’s <strong>Manual</strong><br />

p/n 18247-001, Rev. G