Low Pressure O 2 Source<br />

When selected, this option allows oxygen to be supplied from a low pressure / low flow oxygen<br />

source such as an oxygen concentrator or line mounted flow meter. Oxygen from the low pressure<br />

source is mixed with air inside the ventilator. The O2 percent delivered to the patient is determined by<br />

the O2 inlet flow and the total minute volume and is not regulated by the ventilator. Use the Input O2<br />

Flow chart (page 6-15) to determine the correct O2 flow for the desired FIO2.<br />

• When the Low Pressure O2 Source option is selected and a high O2 pressure source is attached<br />

to the ventilator, an Automatic High O2 Switch Over safety response generates a HIGH O2 PRES<br />

alarm, switches the ventilator to High Pressure O2 Source mode and sets the percentage of<br />

oxygen to be delivered in the gas flow to 21%.<br />

When the Low Pressure O2 Source option is not selected, a high pressure oxygen source is<br />

expected, and oxygen blending is done within the ventilator. The ventilator expects an oxygen source<br />

with a pressure of 40 - 80 PSIG. The O2 percent delivered to the patient is determined by the O2 %<br />

(O2 Flush) setting on the ventilator front panel.<br />

To toggle the state of the Low Pressure O2 Source:<br />

1) Push and hold the Low Pressure O2 Source button for three (3) seconds.<br />

• While the Low Pressure O2 Source button is being held, the associated LED will be flashing<br />

• When the Low Pressure O2 Source is selected, the associated LED will be on continually<br />

While Low Pressure O2 Source is on:<br />

• The O2 Inlet Pressure Low alarm is inactive.<br />

• The O2 Pressure High alarm is set to activate at > 10 PSIG.<br />

• The O2 % (O2 Flush) display will display dimmed dashes and O2 % cannot be set.<br />

• Oxygen inlet flow must be set to obtain the desired oxygen percentage.<br />

WARNING<br />

Inspired Oxygen (FIO2) Concentration – If the patient has a variable respiratory rate, his/her minute<br />

ventilation will fluctuate. If exact concentrations of inspired oxygen (FIO2) are required to be delivered<br />

to the patient, it is recommended that an accurate oxygen analyzer with alarms be used.<br />

AVERTISSEMENT<br />

Concentration d’oxygène inspiré (FIO2) – Si la fréquence respiratoire du patient est variable, sa<br />

ventilation-minute va fluctuer. Lorsqu’une concentration exacte d’oxygène inspiré (FIO2) est<br />

nécessaire pour une transmission au patient, il est recommandé d’utiliser un analyseur de niveau<br />

d’oxygène 2 précis, comportant des alarmes.<br />

Operator’s <strong>Manual</strong> <strong>LTV</strong> ® <strong>1200</strong> Ventilator Page 6-13<br />