Software Version<br />

Use the Software Version option to verify the software version installed in the ventilator. The software<br />

version number is displayed as: VER xx.xx<br />

Usage Meter<br />

Use the Usage Meter to view the time the ventilator has been in use. It is updated every 1/10 th hour<br />

up to 139,000.0 hours and is displayed as: USAGE xxxxxx.x<br />

Communications Setting<br />

The ventilator may be connected to a printer, a graphics monitor, or a modem, or may be set up to<br />

output system diagnostic data. Use the Communications Setting option 49 to select the<br />

communications protocol for data transmission.<br />

• Use the MONITOR setting to communicate with an LTM Graphics Monitor.<br />

To modify the Communications Setting:<br />

1) Push the Select button while COM SETTING is displayed.<br />

2) DATA or the currently selected protocol is displayed.<br />

3) Turn the Set Value knob until the desired protocol is displayed.<br />

4) Push the Select button.<br />

Options: DATA, MONITOR, PRINTER, and MODEM<br />

49 Only DATA and MONITOR (for LTM Graphic Monitors compatible <strong>LTV</strong> ® ventilators) are available at this time.<br />

Page 10-16 <strong>LTV</strong> ® <strong>1200</strong> Ventilator Operator’s <strong>Manual</strong><br />

p/n 18247-001, Rev. G