Tidal Volume<br />

Use the Tidal Volume control to establish the volume of gas which the ventilator will produce and<br />

deliver during Volume Controlled breaths. Flow is delivered in a taper waveform over the set<br />

Inspiratory Time. The peak flow is calculated based on the Tidal Volume and Inspiratory Time with a<br />

maximum flow of 100 lpm and a minimum flow of 10 lpm. Flow is decelerated from the calculated<br />

peak flow to 50% of the calculated peak flow.<br />

While the Tidal Volume is being updated, the Calculated Peak Flow is displayed in the display<br />

window.<br />

To set the Tidal Volume:<br />

1) Push the Tidal Volume button.<br />

2) Change the setting using the Set Value knob.<br />

3) Push the Tidal Volume button again to deselect the setting and accept the new value.<br />

Range: 50 - 2000 ml<br />

NOTE<br />

Be sure that Volume ventilation is selected.<br />

REMARQUE<br />

Assurez-vous que le volume de ventilation est sélectionné.<br />

Page 6-28 <strong>LTV</strong> ® <strong>1200</strong> Ventilator Operator’s <strong>Manual</strong><br />

p/n 18247-001, Rev. G