Chapter 6 - CONTROLS<br />

This section explains how each of the <strong>LTV</strong> ® <strong>1200</strong> ventilator front panel controls work.<br />

Assist/Control - SIMV/CPAP - NPPV<br />

This button toggles between Assist/Control, SIMV/CPAP, and NPPV modes of ventilation. Each<br />

press of the button advances or confirms your selection as follows:<br />

• One press selects the first mode in the series (Assist/Control) and the associated LED<br />

flashes. A second press confirms the selection and the associated LED lights solid.<br />

• If you press the button again, the SIMV/CPAP LED flashes. Pressing again confirms and<br />

causes the SIMV/CPAP LED to light solid green.<br />

• Press once more to select NPPV mode and the NPPV LED flashes. To confirm NPPV mode<br />

you must press again but be aware that the NPPV LED will continue to flash until the IPAP<br />

and EPAP values have been set.<br />

• For more information see Procedure for NPPV Mode Set Up, in Chapter 12 – Operating<br />

Procedure.<br />

NOTE<br />

When Assist/Control is selected, the ventilator will be in Control or Assist/Control mode, depending<br />

on the Sensitivity setting.<br />

• If Sensitivity is set to dashes “- -”, the ventilator will be operating in Control mode.<br />

• If Sensitivity is set to any other value, the ventilator will be operating in Assist/Control mode.<br />

When SIMV/CPAP is selected, the ventilator will be in SIMV or CPAP mode, depending on the<br />

Breath Rate setting.<br />

• If Breath Rate is set to dashes “- -”, the ventilator will be operating in CPAP mode.<br />

• If Breath Rate is set to any other value, the ventilator will be operating in SIMV mode.<br />

REMARQUE<br />

Lorsque Aide / Contrôle est sélectionné, le ventilateur sera en mode Contrôle ou Aide, selon le<br />

réglage de la sensibilité.<br />

• Si la sensibilité est réglée sur Traits « - - », le ventilateur fonctionnera en mode Contrôle.<br />

• Si la sensibilité est réglée sur toute autre valeur, le ventilateur fonctionnera en mode Aide /<br />

Contrôle.<br />

Lorsque le mode SIMV/CPAP est sélectionné, le ventilateur sera en mode SIMV ou CPAP, selon le<br />

réglage du débit respiratoire.<br />

• Si le débit respiratoire est réglé sur Traits « - - », le ventilateur fonctionnera en mode CPAP.<br />

• Si le débit respiratoire est réglé sur toute autre valeur, le ventilateur fonctionnera en mode SIMV.<br />

Operator’s <strong>Manual</strong> <strong>LTV</strong> ® <strong>1200</strong> Ventilator Page 6-1<br />

p/n 18247-001, Rev. G