Cleaning or Replacing the O 2 Inlet Filter<br />

The O2 filter should be cleaned or replaced when it becomes soiled. Failure to do this can affect<br />

ventilator performance.<br />

CAUTION<br />

Oxygen Supply Contamination - The accuracy of the oxygen delivery capabilities of <strong>LTV</strong> ® <strong>1200</strong><br />

ventilator can be compromised by foreign debris contamination in the oxygen supply system. To<br />

reduce the risk of airborne contaminants entering the ventilator, ensure that any oxygen supply<br />

connected to the ventilator is clean, properly filtered 54 and that the ventilator’s O2 Inlet Port Cap is<br />

securely installed on the O2 Inlet Port whenever the ventilator is not connected to an external oxygen<br />

supply.<br />

ATTENTION<br />

Contamination de la réserve d’oxygène - La précision de la capacité d’alimentation en oxygène<br />

des ventilateurs <strong>LTV</strong> ® <strong>1200</strong> peut être compromise par la présence de corps étrangers dans le<br />

système d’alimentation en oxygène. Afin de diminuer le risque de présence d’agents contaminants<br />

atmosphériques dans le ventilateur, assurez-vous que la réserve d’oxygène reliée au ventilateur est<br />

propre et filtrée de manière adéquate 54 , et que le bouchon de l’orifice d’alimentation en oxygène est<br />

correctement installé à chaque fois que le ventilateur n’est pas relié à une source d'oxygène externe.<br />

To clean or replace the O2 Inlet Filter:<br />

1) If a high pressure O2 source is being used, disconnect the high pressure O2 hose from the<br />

oxygen block on the left side of the ventilator.<br />

2) If a low pressure O2 source is being used, disconnect the O2 line from the barbed oxygen<br />

adapter. Unscrew and remove the barbed adapter from the oxygen block on the left side of the<br />

ventilator.<br />

54<br />

In addition to the existing internal O2 Inlet filter, P/N 14313, an External, In-Line Oxygen Filter (P/N 14470) is<br />

available from Pulmonetic Systems.<br />

Page 13-4 <strong>LTV</strong> ® <strong>1200</strong> Ventilator Operator’s <strong>Manual</strong><br />

p/n 18247-001, Rev. G