LOW PRES<br />

When the peak inspiratory pressure for a selected breath is less than the Low Pressure setting, the<br />

LOW PRES alarm is generated. The Low Pressure alarm can be set to apply to all breaths (ALL<br />

BREATHS) or to Volume Control (VC) and Pressure Control (PC) breaths only. (For information on<br />

selecting breath types, see Chapter 10 - Extended Features, Low Peak Pressure Alarm.)<br />

When a LOW PRES alarm occurs:<br />

• The LOW PRES message is flashed in the display window.<br />

• The Low Pressure control display is flashed.<br />

• The audible alarm is sounded.<br />

To reset the LOW PRES alarm:<br />

1) Push the Silence Reset button twice.<br />

WARNING<br />

Patient Circuit Accessories - The use of accessories such as Speaking Valves, Heat-Moisture<br />

Exchangers and Filters create additional patient circuit resistance and in the event of a disconnection,<br />

may impede the generation of a Low Pressure alarm. Ensure that the Low Pressure alarm settings<br />

accommodate these types of accessories when used in combination with patient circuits.<br />

AVERTISSEMENT<br />

Accessoires du circuit du patient - L’utilisation d’accessoires tels que les membranes vocales, les<br />

échangeurs thermohydriques et les filtres, produit une résistance additionnelle dans le circuit de<br />

patient et en cas de débranchement, elle risque d’empêcher la génération de l’alarme de basse<br />

pression. S’assurer que les paramètres de l’alarme de basse pression s’adaptent à ces types<br />