Exhalation Flow Transducer Differential Autozero - Wide<br />

Use this item to view the Exhalation Flow Transducer Differential Autozero - Wide results and<br />

schedule the Exhalation Flow Transducer Differential Autozeros - Wide to be run.<br />

To view the Exhalation Flow Transducer Differential Autozero – Wide results:<br />

1) The previous results, FDw xxxx P, are displayed. The final P indicates the previous zero results<br />

were within the required tolerance and the previous autozero passed. If a final F is displayed, the<br />

previous zero results were outside the required tolerance and the autozero failed. An asterisk<br />

indicates that an autozero is scheduled for the next breath.<br />

2) Turn the Set Value knob to display the EXIT option.<br />

3) Push the Select button.<br />

To schedule the Exhalation Flow Transducer Differential Autozero - Wide:<br />

1) The previous results, FDw xxxx P, are displayed.<br />

2) Push the Select button. An asterisk appears, the pass / fail indicator is removed from the display<br />

and the autozero test is scheduled for the next breath.<br />

3) After the autozero is run on the next breath, the new autozero value and the pass / fail indicator<br />

are displayed.<br />

If the autozero fails, it will be automatically rescheduled for the next breath.<br />

Operator’s <strong>Manual</strong> <strong>LTV</strong> ® <strong>1200</strong> Ventilator Page 10-31<br />