28.02.2013 Views

LTV 1200 Operator's Manual - CareFusion Emergency Preparedness

LTV 1200 Operator's Manual - CareFusion Emergency Preparedness

LTV 1200 Operator's Manual - CareFusion Emergency Preparedness

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Appendix B - SET UP / MAINTENANCE<br />

Recommended Maintenance Schedule<br />

The <strong>LTV</strong> ® <strong>1200</strong> ventilator is designed to operate for extended periods of time with minimal routine<br />

maintenance. The following periodic maintenance is recommended:<br />

Hours of Service 66 Maintenance Required<br />

Prior to initial use • Charge the Internal Battery by plugging the ventilator into an AC power<br />

source for 24 hours.<br />

• Setup the ventilator/accessories per Appendix C - Installation and<br />

Checkout.<br />

• Check the ventilator for proper operation per Appendix C - Installation and<br />

Checkout<br />

While in storage,<br />

every two months<br />

• Recharge the Internal Battery by plugging the ventilator into an AC power<br />

source for 24 hours 67 .<br />

Daily • Check the Inlet Filter, clean if necessary.<br />

• Check the Fan Filter, clean if necessary.<br />

If in use, a minimum<br />

of once a month<br />

Every 10,000 hours<br />

or two years 68<br />

Every 30,000 hours<br />

or six years,<br />

whichever comes<br />

first 66<br />

• Check the ventilator per Chapter 11 - Ventilator Checkout Tests. While<br />

the ventilator is off-patient, perform the Power (external) disconnect test*<br />

• Verify Vte or VE monitor *<br />

• Verify airway pressure or PIP monitor *<br />

• Verify delivered O2 concentration if not using an oxygen analyzer<br />

continuously, See Appendix A - Ventilator Specifications for accuracy<br />

tolerances.<br />

* Use existing patient settings or example settings as shown in Appendix C<br />

- Installation and Checkout<br />

• Replace the Internal Battery 69 only with Pulmonetic Systems battery P/N<br />

18608-001 70<br />

• Calibrate the Transducers.<br />

• Replace the Motor Board.<br />

• Clean or replace the Interior Air Inlet Filter.<br />

• Clean or replace the O2 Inlet Filter.<br />

• Replace the Turbine Manifold Assembly.<br />

• Replace the Solenoid Manifold.<br />

• Replace the Flow Valve.<br />

• Replace the Rotary Switch Assembly.<br />

• Replace the O2 Blender.<br />

• Replace the Fan Assembly.<br />

• Replace all Silicone Tubing.<br />

• Check the Thermo Pads for compression and replace if necessary.<br />

66 To check the number of hours the ventilator has been in service, see Chapter 10 - Extended Features, Usage Meter.<br />

67 If the battery is deeply discharged, it may take several charge and discharge cycles before it can be fully charged.<br />

68 10,000 hour, two year and/or 30,000 hour, six year Extended Maintenance and ventilator repair must be performed by a<br />

Pulmonetic Systems factory trained service technician.<br />

69 Replacement at 10,000 hours or 2 years is based on normal use of up to 200 charge cycles. The battery may need to be<br />

replaced more frequently if it is being charged more often. The battery should also be replaced any time it fails to reach a full<br />

charge, or if the ventilator runs for less than ½ hour on a fully charged battery.<br />

70 The <strong>LTV</strong> ® Internal Battery (P/N 18608-001) is contained in <strong>LTV</strong> ® Internal Battery Replacement Kit, P/N 18634-001.<br />

Operator’s <strong>Manual</strong> <strong>LTV</strong> ® <strong>1200</strong> Ventilator Page B-1<br />

p/n 18247-001, Rev. G

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!