Peak<br />

Flow<br />

•<br />

V<br />

Set<br />

Pressure<br />

Paw<br />

Pressure Control Flow Termination<br />

Use the Pressure Control Flow Termination option to enable or disable flow termination for Pressure<br />

Control breaths.<br />

When this option is ON, Pressure Control breaths are cycled at the set percentage of the peak flow if<br />

it is reached before the set Inspiratory Time elapses. The percentage of peak flow is set in the<br />

Variable Flow Termination option.<br />

When this option is OFF, Pressure Control breaths are cycled when the set Inspiratory Time is<br />

reached.<br />

Set pressure is<br />

maintained and<br />

inspiratory flow<br />

goes to zero<br />

Set Insp Time<br />

To modify the Pressure Control Flow Termination setting:<br />

1) Push the Select button while PC FLOW TERM is displayed.<br />

2) PC FLOW ON or PC FLOW OFF is displayed.<br />

3) Turn the Set Value knob until the desired state is displayed.<br />

4) Push the Select button.<br />

Options: ON or OFF<br />

Breath<br />

terminates<br />

at set Insp<br />

Time<br />

PC FLOW TERM set to OFF<br />

Pressure Control Breath terminates normally<br />

Operator’s <strong>Manual</strong> <strong>LTV</strong> ® <strong>1200</strong> Ventilator Page 10-11<br />

p/n 18247-001, Rev. G<br />

Peak<br />

Flow<br />

•<br />

V<br />

Set<br />

Pressure<br />

Paw<br />

Breath terminates<br />

at percentage of<br />

peak flow<br />

Set Insp Time<br />

Set Percentage of<br />

Peak Flow<br />

PC FLOW TERM set to ON<br />

Pressure Control Breath terminates at the same<br />

Percentage of Peak Flow as Pressure Support breaths