Battery Level<br />

The Battery Level indicator shows the level of available internal battery power while running from the<br />

internal battery. When the ventilator is running from an external power source, the Battery Level<br />

indicator is off. When running from the internal battery at the nominal settings shown below, the<br />

indicator shows the following levels:<br />

Alarm<br />

LED<br />

Color Battery Level Approximate Battery Time<br />

(Total time: 60 mins)<br />

-- Green Internal battery level is acceptable 45 minutes<br />

BAT LOW Amber Internal battery level is low 10 minutes<br />

BAT EMPTY Red Internal battery level is critically low 5 minutes<br />

Mode<br />

Nominal Settings<br />

Assist/Control, Volume Peep 5 cmH2O<br />

Breath Rate 15 bpm 02 % 21%<br />

Tidal Volume 800 ml Lung Compliance 50 ml/cmH2O<br />

Inspiratory Time 1.5 sec ET Tube Resistance 5.87 cmH2O/L/s<br />

Sensitivity 2 lpm Battery Temperature 25° C<br />

When an <strong>LTV</strong> ® <strong>1200</strong> ventilator is operated on its internal battery to the point that the internal battery is<br />

completely depleted, the ventilator will shut down. If the ventilator remains in this state, the internal<br />

battery may recharge slightly within a few seconds / minutes and cause the ventilator to automatically<br />

restart and operate for a short period of time. This cycle may repeat several times, depending on the<br />

condition of the internal battery.<br />

Page 7-2 <strong>LTV</strong> ® <strong>1200</strong> Ventilator Operator’s <strong>Manual</strong><br />

p/n 18247-001, Rev. G