WARNING<br />

• Put the MR Conditional <strong>LTV</strong> ® <strong>1200</strong> System outside the 100 Gauss field line (at least 9ft / 2.74m<br />

from the bore opening) and in a position where the operator can always see the <strong>LTV</strong> ® <strong>1200</strong>.<br />

Always lock the two (2) locking caster wheels on the ventilator floor stand when used in the MR<br />

environment.<br />

• Use only MR Conditional non-magnetic (e.g., aluminum) oxygen cylinders, and MR Conditional<br />

oxygen regulators.<br />

AVERTISSEMENT<br />

• Placez le système à RM conditionnelle <strong>LTV</strong> ® <strong>1200</strong> à l'extérieur de la ligne de force de 100<br />

Gauss (au moins 2,74 m/9 pi de l'ouverture d'alésage) et dans une position où l'opérateur peut<br />

toujours voir le <strong>LTV</strong> ® <strong>1200</strong>. Verrouillez toujours les deux (2) roulettes verrouilables sur le support<br />

de plancher du ventilateur en utilisant dans l'environnement de RM.<br />

• Utilisez seulement des régulateurs d'oxygène et des bobonnes d'oxygène non-magnétique<br />

(comme en aluminium) à RM conditionnelle.<br />

Page 16-4 <strong>LTV</strong> ® <strong>1200</strong> Ventilator Operator’s <strong>Manual</strong><br />