O 2 Cylinder Duration (cont.)<br />

4) Turn the Set Value knob until CYL PRES is displayed, push the Select button and xxx psi is<br />

displayed.<br />

5) Turn the Set Value knob until the applicable cylinder pressure is displayed, push the Select<br />

button and the cylinder pressure is set.<br />

• Range: 100 - 2300 psi, in increments of 25, or<br />

• Range: 5 - 150 bar, in increments of 1 (if the selected language uses the bar unit of<br />

measurement)<br />

• This setting is not retained by the ventilator through shut downs and power ups, will be reset<br />

to the factory set default value if the Language setting is changed, and will need to be<br />

reviewed/reset by the operator each time the O2 Cylinder Duration option is used.<br />

• After changing this, or any ventilation setting, wait approximately 20 seconds before selecting<br />

CALCULATE, to allow the ventilator to monitor the oxygen flow that will be used in the<br />

calculation and display of the remaining usable time of the external O2 cylinder.<br />

6) Turn the Set Value knob until CALCULATE is displayed and push the Select button.<br />

• To obtain an accurate duration time estimate, the current cylinder pressure must be entered<br />

prior to each calculation.<br />

• After changing any ventilation setting, wait approximately 20 seconds before selecting<br />

CALCULATE, to allow the ventilator to monitor the oxygen flow that will be used in the<br />

calculation and display of the remaining usable time of the external O2 cylinder.<br />

• When CALCULATE is selected, the ventilator uses the current ventilation values and settings<br />

to calculate the remaining usable time of the external O2 cylinder specified and displays O2<br />

DUR hh:mm (O2 duration in hours and minutes) for 60 seconds or until the message is<br />

acknowledged by pushing the Select or Control Lock button, or by rotating the Set Value<br />

knob on the front panel.<br />

• Breath to breath variations may cause slightly different results in consecutive calculations.<br />

Page 10-22 <strong>LTV</strong> ® <strong>1200</strong> Ventilator Operator’s <strong>Manual</strong><br />

p/n 18247-001, Rev. G