PEEP Control<br />

The PEEP control establishes the Positive End Expiratory Pressure.<br />

To set the PEEP:<br />

1) Push the PEEP control button.<br />

2) Change the setting using the Set Value knob.<br />

Range: 0 – 20 cmH2O<br />

The <strong>LTV</strong> ® <strong>1200</strong> has PEEP Compensated Pressure Support and Pressure Control.<br />

On the left waveform (breath), the ventilator is set to deliver Pressure Support (or Pressure Control)<br />

of 20 cmH2O with no PEEP. The “span” or “delta” is 20 cmH2O.<br />

On the right waveform, the vent is set to deliver Pressure Support (or Pressure Control) of 20 cmH2O.<br />

Note that 5 cmH2O of PEEP have been added. This ventilator “compensates” for the added PEEP by<br />

automatically increasing the Pressure Support (or Pressure Control) in order to maintain the same<br />

“span” or “delta” of 20 cmH2O.<br />

To view the measured PEEP in the display window:<br />

1) Use the Select button to display the measured PEEP in the display window.<br />

PEEP Control in NPPV Mode<br />

When NPPV mode is selected, the PEEP control is used to set the EPAP value.<br />

Operator’s <strong>Manual</strong> <strong>LTV</strong> ® <strong>1200</strong> Ventilator Page 6-21<br />