NPPV<br />

The NPPV 30 indicator LED is lit when NPPV mode is selected. For more information on NPPV mode,<br />

see NPPV Mode in Chapter 4 - Ventilation Modes and Procedure for NPPV Mode Set Up in Chapter<br />

12 – Operating Procedure.<br />

Patient Effort<br />

This LED is lit briefly each time a patient trigger is detected. See Chapter 6 - Controls, Sensitivity for<br />

more information on patient triggers.<br />

Vent Inop<br />

The Vent Inop LED is lit any time the ventilator is in the Inop state. This occurs when:<br />

• The ventilator is put into Standby using the On/Standby button.<br />

• The ventilator power sources, both external and internal, are insufficient to operate the ventilator.<br />

• A Vent Inop alarm sounds.<br />

An audible alarm sounds continuously when the ventilator enters the Vent Inop state, and may be<br />

silenced by pushing the Silence Reset button.<br />

While in the Vent Inop state, the ventilator is set to a safe state, allowing the patient to breathe<br />

spontaneously from room air.<br />

30 Non-Invasive Positive Pressure Ventilation<br />

Page 7-6 <strong>LTV</strong> ® <strong>1200</strong> Ventilator Operator’s <strong>Manual</strong><br />

p/n 18247-001, Rev. G