MicroStation System, MicroLog Version 4.2 - DTU Systems Biology ...
MicroStation System, MicroLog Version 4.2 - DTU Systems Biology ...
MicroStation System, MicroLog Version 4.2 - DTU Systems Biology ...
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Inoculating MicroPlates<br />
Troubleshooting<br />
Symptom Cause Solution<br />
Pipetting problems Wells inoculated unevenly Dispense first aliquot from pipettor back into reservoir.<br />
Depress lever smoothly and completely.<br />
Make sure tips are lined up evenly and tightly seated.<br />
Avoid pipetting or trapping air bubbles.<br />
Visually check wells after inoculating.<br />
Incorrect volumes used<br />
Incubating MicroPlates<br />
100 µl for anaerobic bacteria, yeasts and fungal.<br />
150 µl for aerobic bacteria.<br />
Symptom Cause Solution<br />
Poor growth or poor<br />
pattern formation in<br />
MicroPlates<br />
Wells at the four<br />
corners or edges of<br />
the MicroPlate have<br />
lower liquid levels<br />
after incubation<br />
Wrong incubation<br />
conditions<br />
Wrong MicroPlate<br />
Organism not in Database<br />
Incubator is too dry<br />
Filamentous fungi not<br />
incubated in a closed<br />
container<br />
Incubate according to specified temperature and<br />
conditions. See Section 4 and Appendices.<br />
Incubate the MicroPlate at the same temperature as<br />
the growth plate.<br />
Make sure incubator humidity is sufficient.<br />
Add a pan of water to humidify incubator, or incubate<br />
MicroPlates in a plastic container humidified with<br />
moist paper towels.<br />
Make sure you have the FF MicroPlates in a closed<br />
container or in plastic bags on trays. Do not add<br />
additional moisture when incubating filamentous<br />
fungi.<br />
<strong>MicroStation</strong> <strong>System</strong>/<strong>MicroLog</strong> User Guide Apr 07 Section 11 � Page 5