<strong>MOSAIQ</strong> User’s <strong>Guide</strong><br />

3. Type the block required for treatment in the Block field.<br />

4. Type the bolus required for treatment in the Bolus field.<br />

If you are using barcode verification of accessory devices, type the<br />

word ‘Custom’.<br />

Manually Creating Geometric and Couch Settings<br />

Most of the time, you transfer treatment field geometric settings from the treatment planning<br />

system that you use. You can also add the geometric and couch settings manually if<br />

necessary.<br />

1. In the Gantry/Collimator group of the Treatment Field Definition dialog box, type the<br />

gantry angle in the Gantry Angle field.<br />

2. Type the collimator angle in the Collimator Angle field.<br />

3. Type the width and length of the X jaws in the Field Size X field.<br />

4. Type the width and length of the Y jaws in the Field Size Y field.<br />

5. If the X jaws are asymmetrical, click the Asy check box adjacent to the X jaw fields, then<br />

type the width/length of each X jaw in the Jaw X1 and Jaw X2 fields.<br />

6. If the Y jaws are asymmetrical, click the Asy check box adjacent to the Y jaw fields, then<br />

type the width/length of each Y jaw in the Jaw Y1 and Jaw Y2 fields.<br />

7. Type the couch settings in the Couch group, such as the Angle field.<br />

Manually Creating Portal Image Information<br />

You can specify MU and other parameters specifically for portal images.<br />

1. In the Portal Image group of the Treatment Field Definition dialog box, type the planned<br />

portal image monitor units in the Planned MU field.<br />

You must type a MU value less than the maximum portal image MU characterized for<br />

the selected machine. If you type a value greater than the maximum portal image MU, an<br />

error message appears.<br />

2. Type the open field portal image monitor units in the Open MU field.<br />

You can use the Dose Coeff field if your facility adds the dose from<br />

portal images to the total dose.<br />

3. Type the incremental number of centimeters the collimator field size opens from the<br />

treatment field size for an open field portal image exposure in the Delta field.<br />

The EPID and SID check boxes are required fields for the Varian 4D<br />

and Siemens Coherence machines.<br />

94 Copyright 2010, Elekta Impac Software<br />

LUGMSQ0002/4.0/23-Sep-2010