<strong>MOSAIQ</strong> User’s <strong>Guide</strong><br />

Appendix B: Warnings and Cautions in<br />

<strong>MOSAIQ</strong><br />

<strong>MOSAIQ</strong> Treatment Warnings<br />

The SEQUENCER verify and record system is primarily a quality assurance and recording<br />

tool. The purpose of SEQUENCER is to help therapists deliver treatments to patients.<br />

SEQUENCER checks setup parameters and records actual treatment data. SEQUENCER is<br />

not a substitute for the many checks that a therapist or the linac makes before and during<br />

treatment to make sure that the patient is setup and treated according to plan. During the<br />

course of treatment, SEQUENCER shows appropriate warning messages that alert the user<br />

to situations in which improper use of the system could result in potential safety hazards.<br />

Whenever you see a warning message, carefully read the description of the potential safety<br />

hazard and the actions to take to avoid compromising the quality of the treatment.<br />

Remember the following when reading SEQUENCER Messages:<br />

SEQUENCER only verifies field parameters within the tolerances established by the<br />

department and selected for the field. Do NOT rely on SEQUENCER to verify the<br />

patient and linac setup position. Check the treatment setup against the parameters in<br />

the Treatment Field Definition.<br />

SEQUENCER receives the actual position and console setting information from the<br />

electronic readouts on the treatment machine. If the electronic readouts are incorrect,<br />

then the information SEQUENCER receives is also incorrect.<br />

Never override a treatment field parameter before you know why the parameter<br />

must be overridden. An out-of-tolerance parameter does not necessarily mean that<br />

the treatment setup is incorrect.<br />

Always verify the treatment setup using an alternative method or methods in<br />

addition to SEQUENCER.<br />

After beaming on the linac, SEQUENCER cannot turn the beam off. You must<br />

closely monitor the linac and the patient during treatment and should not look at the<br />

SEQUENCER screen.<br />