CHAPTER 7: Image Management-ViewStation & Photos and Diagrams<br />

3. Click the image to use in the Stereoscopic Image Selection dialog box. If you click<br />

Cancel, the Image Registration window opens for the initially selected image.<br />

4. In the Image Registration window, reticle scale the verification images, then click the<br />

check box for registration method you wish to use.<br />

You can use different methods for each image except for grayscale matching, which you<br />

must use on both images. Refer to Registering Images on page 197.<br />

5. After you have completed the tasks for your registration method in the Image<br />

Registration window, click Calculate Shifts to see the fused image pair.<br />

The stereoscopic registration calculates offset information in 3 dimensions, which is<br />

shown on the Isocenter Shifts tab of the Isocenter Shifts window. The magnitude shown<br />

is the linear distance of the correction in 3D space (the diagonal of the 3 correctional<br />

directions).<br />

6. Click OK in the Isocenter Shifts window when you have the offset you want on the<br />

Isocenter Shifts tab.<br />

If you are doing offline image registration using Setup Intelligence,<br />

open the image pair from the Image Work List dialog box, then<br />

follow steps 2-6 above. Refer to Managing the Review Process – for<br />

Physicians on page 182 and Managing the Review Process – for<br />

Therapists on page 187 for instructions on how to use the Image<br />

Work List dialog box.<br />

The Active Check box in the Image Review Window<br />

In the Image Review window, the Active check box on the Isocenter Offset tab is selected<br />

when the Session Offset is included in the Localization Trend Review. To include a zero<br />

value in analysis, you must manually enter 0 in each orientation and save (zero ≠ null). Refer<br />

to Figure 55.<br />

Figure 55: Active Check box on Isocenter Shifts Tab<br />

Copyright 2010, Elekta Impac Software 209<br />