<strong>MOSAIQ</strong> User’s <strong>Guide</strong><br />

Auto-Setup You should only use Auto Setup as an approximate method to<br />

position the linac to treat a patient. You cannot rely solely on<br />

the electronic readouts to position the linac. You must verify<br />

the linac position with an alternative method or methods.<br />

Improper use of the auto setup feature could result in serious<br />

injury to the patient. You should be in the treatment room, and<br />

constantly monitor the treatment unit while you use Auto<br />

Setup. You must be thoroughly familiar with the emergency<br />

procedures to stop motions (disengaging motion enable<br />

switches, pressing emergency off, etc.) before you use the autosetup<br />

feature. If a collision condition appears likely, it is your<br />

responsibility to stop motion and disengage the motion enable<br />

switch or take other recommended action to stop motion as<br />

recommended by the linac manufacturer (press the emergency<br />

off).<br />

Geometric<br />

Tolerances<br />

SEQUENCER only verifies geometric setup parameters to<br />

within the specified tolerances. Do NOT rely on SEQUENCER<br />

to verify the patient and linac setup position. Check the<br />

treatment setup using the parameters specified in the Treatment<br />

Field Definition dialog box.<br />

Treatment Setup You cannot use SEQUENCER as the sole method to verify and<br />

record the treatment setup. You must use an alternative method<br />

or methods to verify and record the treatment setup in addition<br />

to SEQUENCER.<br />

Recording<br />

Treatments<br />

Always review all information in the Record Treatment dialog<br />

box before saving it. Make sure that the correct patient and<br />

treatment field appears. Make sure that the correct number of<br />

monitor units appears as compared to the electronic and/or<br />

mechanical monitor units readout on the linac. If the monitor<br />

units value is not the same in SEQUENCER and the linac, then<br />

record the monitor units value on the linac as a Note and<br />

contact your department physicist immediately before you<br />

continue the treatment.<br />