CHAPTER 7: Image Management-ViewStation & Photos and Diagrams<br />

3. Click adjacent to the Patient Orientation field, then click the applicable patient<br />

orientation setting.<br />

Figure 36 shows the drop list of patient orientation settings in the system.<br />

Figure 36: Patient Orientation Settings<br />

For safety purposes, there is no default patient orientation.<br />

Configuring Treatment Field Definition Data<br />

Before using Image Registration, you must enter the gantry and couch angles in the<br />

Treatment Field Definition dialog box<br />

With regard to the gantry angle, 2D Image Registration is only available for images at<br />

cardinal angles. Cardinal angles are 0°, 90°, 180°, and 270°.<br />

With regard to the couch angle, the couch must be perpendicular to the gantry. This angle<br />

will be either 0.0° or 180° depending on your machine characterization.<br />

Figure 37 shows the Gantry/Collimator and Couch sections in the Treatment Field Definition<br />

dialog boxes.<br />

Figure 37: Gantry and Couch Angles in the Treatment Field Definition<br />