<strong>MOSAIQ</strong> User’s <strong>Guide</strong><br />

SEQUENCER<br />

Intended Use<br />

The intended use of SEQUENCER (generically referred to as a verification system) is to assist<br />

in the process of patient treatment on a radiotherapy treatment machine. SEQUENCER<br />

provides the capability to notify the users of actions that need to take place prior to<br />

treatment, to display reference images for se up purposes, and to auto setup the machine to<br />

predefined settings. The actual settings are read from the treatment machine, via the<br />

machine’s communication interface, and compared to the predefined values. If a mismatch<br />

occurs between the planned values and the actual machine settings, treatment is prohibited.<br />

At appropriate points during the treatment, the actual delivered values are recorded to<br />

provide treatment tracking.<br />

Treatments are initiated by the operator, via the machine control system, only after the user<br />

has verified that the treatment setup is correct. SEQUENCER does not bypass any checks put<br />

in place by the machine’s control system.<br />

The intended benefit of a computer based verification system is to eliminate the errors which<br />

occur in treatment setup when the only means of verification is the user’s manual comparison<br />

of the handwritten chart against the machine settings and to record actual treatment<br />

parameters, directly from the machine, again eliminating the errors which occur from manual<br />

recording and calculation. Thus, a verification system is an adjunct to the normal treatment<br />

regimen intended to enhance total treatment quality.<br />

Indications for Use<br />

SEQUENCER is to be used to facilitate the delivery of defined radiotherapy treatment plans.<br />

SEQUENCER verifies the settings on a radiotherapy treatment machine prior to treatment<br />

and records the actual parameters after treatment. SEQUENCER can be used whenever<br />

radiotherapy treatment is prescribed.<br />

292 Copyright 2010, Elekta Impac Software<br />

LUGMSQ0002/4.0/23-Sep-2010